# Major Recall Alert: Getinge Washer/Disinfector Recalled Due to Potential Overheating Risk
## Why This Recall is Important
On **August 21, 2025**, the **Getinge Washer/Disinfector 88-5 model** was officially recalled in Canada due to a critical safety issue. The defect stems from a potential overheating risk caused by improper **circulation pressure calibration** during operation. If the calibration was skipped or performed incorrectly, and the low-pressure alarm fails to trigger, the risk of overheating becomes significantly heightened.
This recall is specific to the **electrical heated 88-5 washer disinfectors**, as their construction differs from steam-heated versions, which do not present the same overheating risks. The **recall aims to protect user safety**, especially in healthcare settings where these devices are critical for ensuring sterilization of medical instruments. By addressing deviations in pressure levels during standby mode, the recall seeks to mitigate the possibility of malfunctions that could compromise clinical operations and patient safety.
## Details of the Recall
Here’s what you need to know about this recall:
- **Brand**: Getinge Washer/Disinfector
- **Product Model**: Electrical heated 88-5 Washer Disinfectors
- **Recall Issue**: Potential overheating due to calibration errors during circulation pressure testing or a failure of the low-pressure alarm to trigger.
- **Exclusion**: Steam-heated models are NOT affected, as their construction prevents overheating of heating elements.
- **Reason for Action**: To verify the circulation pressure value in standby mode is at **0.0 kPa**. Any deviation requires recalibration by a **qualified technician** to ensure safe and reliable operation.
- **Recall Start Date**: **August 21, 2025**
If unaddressed, this issue could pose a **significant fire hazard**, lead to equipment failure, or compromise the safety of sterilization processes within clinical environments.
For up-to-date information, visit the **official recall page**: [Getinge Washer/Disinfector Recall](https://recalls-rappels.canada.ca/en/alert-recall/getinge-washerdisinfector).
## What You Should Do
If you own or operate the **Getinge 88-5 Electrical Heated Washer Disinfector**, follow these steps immediately:
- **Inspect the Device**: Confirm whether your unit belongs to the identified affected models. Check the product label or reach out to Getinge’s **customer service team** for product verification.
- **Contact a Qualified Technician**: A certified technician should inspect the **circulation pressure calibration in standby mode**. Ensure the value reads **0.0 kPa**. Devices showing deviations must be recalibrated to prevent safety risks.
- **Pause Usage (If Necessary)**: If recalibration isn’t possible immediately, discontinue use to avoid potential overheating.
- **Notify Your Team**: Alert staff responsible for operating the equipment to ensure safety protocols are followed until the issue is resolved.
If you have further questions or concerns, contact Getinge’s **support team** or refer to the official recall announcement linked above.
## Stay Safe – Get Instant Recall Alerts
Recalls like this highlight the importance of staying informed and proactive about issues impacting critical devices. Don’t wait—stay ahead of potential risks with instant alerts sent directly to your phone.
**Download our recall alerts app now** to keep your workplace, home, and family safe! Stay informed about the latest safety notices across healthcare, consumer goods, automotive, and more.
[Download the Recall Alerts App Today!](#)
---
The **Getinge Washer/Disinfector recall** underscores the risks associated with improper calibration and why regular maintenance is vital. Ensure compliance by acting promptly to protect users, staff, and patients from avoidable hazards. For more details, check out the [official recall notice](https://recalls-rappels.canada.ca/en/alert-recall/getinge-washerdisinfector).