# Major Recall Alert: Automated Impella Controller Experiences Delayed "Purge System Blocked" Alarm – Update Software to Stay Safe
Abiomed has issued an important recall for the **Automated Impella Controller (AIC)** with software version **V8.2.3**, particularly when used with the **Impella 5.5 with Smart Assist set** (product code 0550-0004). The recall highlights a critical safety issue: a **delayed "purge system blocked" alarm**, which could pose serious risks to patients. While the issue has been resolved in software versions post-**V10.1**, users still operating with the older version are urged to **update their AIC software** immediately.
This blog post will break down the details of the recall, explain why it is essential to act, and provide steps to keep patients and healthcare facilities safe.
## Why This Recall is Important
Medical devices like the Automated Impella Controller are designed to assist patients during critical times, such as during heart procedures. A delay in the **"purge system blocked" alarm** can lead to significant complications, potentially compromising patient outcomes. This recall ensures that healthcare providers and facilities address the problem as quickly as possible to prevent any risks related to delayed alerts.
Key Points to Note:
- The **recall applies specifically** to the AIC with software version **V8.2.3** when paired with the **Impella 5.5 with Smart Assist set** (product code 0550-0004).
- Abiomed confirms that **software versions V10.1 and above** have resolved this delayed alarm issue.
- To ensure patient safety, it is **critical to update the software** on all affected devices or stop using the AIC with Impella 5.5 components until updated.
Staying proactive with medical device recalls not only protects patients but also ensures the integrity of healthcare standards.
## Details of the Recall
Abiomed's recall impacts the following products:
- **Product Name**: Automated Impella Controller (AIC)
- **Affected Software Version**: V8.2.3
- **Issue**: Delayed "purge system blocked" alarm when used with Impella 5.5 with Smart Assist set (product code 0550-0004)
- **Resolved in Software Versions**: V10.1 or later
The delayed purge alarm could inhibit timely responses to complications, which might have serious consequences during use. For this reason, Abiomed strongly advises against using the affected devices until the software is updated to **V10.1 or above.**
For more details, you can access the official recall announcement here: [Automated Impella Controller Recall](https://recalls-rappels.canada.ca/en/alert-recall/automated-impella-controller).
## What You Should Do
Here’s a step-by-step guide to address the recall:
1. **Identify Your Device**: Determine if you are using an Automated Impella Controller running software version **V8.2.3**.
2. **Check for Compatibility**: Verify if the AIC is being used with the Impella 5.5 with Smart Assist set (product code 0550-0004).
3. **Update the Software**: Update the AIC software to version **V10.1 or later** to resolve the delayed alarm issue.
4. **Follow Abiomed’s Recommendations**: Do not use the Automated Impella Controller with affected devices until the software is updated.
5. **Communicate with Your Team**: Ensure that all relevant personnel within your facility are aware of the recall and take the necessary steps to minimize risks.
Staying vigilant and up-to-date on software improvements can prevent critical problems and ensure optimal device performance.
## Stay Safe – Get Instant Recall Alerts
Healthcare providers and facilities must prioritize patient safety by staying informed about medical device recalls. To ensure you never miss an important update, consider subscribing to automated recall notices or downloading tools that provide instant alerts for product recalls and updates.
For further information and official details about this recall, please visit the Canadian government’s recall advisory page: [Automated Impella Controller Recall](https://recalls-rappels.canada.ca/en/alert-recall/automated-impella-controller).
By taking prompt action, users of the Automated Impella Controller can safeguard their patients while ensuring compliance with Abiomed’s updated guidelines. Don't delay—**update your AIC software to V10.1 or above today** to maintain the highest safety standards in your facility.